Functions:
- Automatic infusion
- KVO (Keep Vein Open)
It will give audible and visible alarm and then switch to KVO mode after finish infusing the preset volume
- Audible and visible alarm
The pump has OCCLUSION, AIR IN LINE, DOOR OPEN, FINISH, LOW BATT, INTERRUPTION, ABNORMAL and DELAY OPERATION alarm functions
- AC/DC
When the pump stops, purge function is ended to purge air out of line. Double click bolus button to purge. Contrarily, it infuses fast
- Flow rate setting
Set the flow rate between: Drop/min, ml/h and time infusion volume
- Self-check after startup
When power on, the pump executes a self-check. LCD display and all lamps will be lighten if it works normally. The corresponding lamp will be off if any part doesn't work
- Accumulated volume display and clear
- Memory function
Memory can keep the parameters which you set up last time over 8 years
- Flow rate calibration
In order to make infusion accurate, calibrate the infusion set according to actual situation
Specifications:
Flow Rate
|
1ml/h~1300ml/h
More than 1000ml/h, 1ml/h step
Less than 1000ml/h, 0.1ml/h step
|
Standard IV SET
|
±5%
|
Preset infusion volume
|
1ml~9999ml
More than 1000ml, 1ml step
Less than 1000ml, 0.1ml step
|
Occlusion sensitivity
|
Optional choose from higher to lower
|
KVO rate
|
1~5ml/h (adjustable)
|
Power supply
|
110~220+10%, 50~60Hz+1Hz
|
Power consumption
|
Less than 20VA
|
Battery Life
|
More than 3 hours continuously at 30ml/h with fully charged battery
|
Work Condition
|
Circumstance Temperature: 10 ~40℃
infusion precision will decrease when temperature ≤18℃
Relative Humidity: 30~93%
Atmospheric Pressure: 86~106KPa
|
Pump dimension (mm)
|
233x146x269
|
Pump weight
|
About 3kg
|
Applicable scope
|
Fit for mainline, small volume infusion, quick infusion
|
|